Z-1227-2020Class II
Coopervision Vertex Toric /2 Clear Toric- Power -2.50 -0.75 /90 Soft (hydrophilic) Contact Lens (extended...
CooperVision, Inc. / initiated 2020-01-13 / Terminated / root cause: manufacturing process
CooperVision Inc. began a recall of Coopervision Vertex Toric /2 Clear Toric- Power -2.50 -0.75 /90 Soft (hydrophilic) Contact Lens (extended wear) on . FDA classified it as Class II and gives the reason as "Incorrect power of lenses than the expected". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1227-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CooperVision, Inc.
- Product
- Coopervision Vertex Toric /2 Clear Toric- Power -2.50 -0.75 /90 Soft (hydrophilic) Contact Lens (extended wear)
- Product code
- LPM Lenses, Soft Contact, Extended Wear
- Reason for recall
- Incorrect power of lenses than the expected
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- CooperVision issued Device Correction letter dated 1/20/19, stating reason for recall, health risk and action to take: Examine your inventory for the impacted lot numbers on the enclosed Ship Report. " Stop any further distribution of the specified lots of Vertex Toric and 2 Clear Toric lenses as appropriate. " Please discard product and contact CooperVision, Customer Service department at 1-800-340-2020 to receive credit and/or complimentary replacement product. Questions contact CooperVision Customer Service at 1-800-341-2020, 9:00 AM 7:00 PM ET, Monday Friday.
- Quantity in commerce
- 216 units
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Lot Numbers: 15627000009010 15627050004010 16656000008010
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1227-2020 | Class II | Terminated | Voluntary: Firm initiated | ||