Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1327-2020Class II

OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565

The Binding Site Group Ltd. / initiated 2019-07-29 / Terminated / root cause: design

The Binding Site Group, Ltd. began a recall of OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565 on . FDA classified it as Class II and gives the reason as "There is a potential that users may experience difficulty validating the calibration of the kit. This is indicated by kit control values reporting outside the specified ranges. This could result in potential false...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1327-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
The Binding Site Group Ltd.
Product
OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565
Product code
CFN Method, Nephelometric, Immunoglobulins (G, A, M)
Reason for recall
There is a potential that users may experience difficulty validating the calibration of the kit. This is indicated by kit control values reporting outside the specified ranges. This could result in potential false patient results which could cause a delay until repeat testing is performed.
Root cause tag
design
FDA root cause
Component design/selection
Action
On July 29, 2019, the firm issued a "Field Safety Notice" to all consignees. The firm requested consignees do the following: 1. For kit NK012.0PT.A lot 422251, please load reagents on-board the analyser and leave overnight, or alternatively for a minimum of 18 hours, prior to running the calibration and validating the curve. The Optilite analyser holds the assay reagent (R012.0PT lot 446963) in a temperature controlled reagent carousel; please leave the reagents on-board the analyser until empty (following stability advice in the insert Section 6). 2. Customers are advised that once the calibration is established and validated,the reagents must remain on the instrument until the reagent is consumed including when re-running the calibration. Investigations on NK012.0PT.A lot 422251 have shown that inversion of the reagent vials prior to re-calibration can contribute to kit control values reporting outside the specified ranges. 3. Control and calibrator vials are held in the sample handling area of the analyser. When not in use, the control and calibrator vials should be removed from the instrument, be capped to avoid evaporation and be stored at 2-sc in the refrigerator. 4. To validate the calibration curve, the control materials must be tested a minimum of once a day and should return results within 15% of the concentration(s) as stated on the accompanying QC certificate. If a control measurement is out of range when assayed with a stored curve the assay must be recalibrated. If on recalibration the control values measured with the new curve are still out of range, the instrument should be checked before repeating the assay 5. In our laboratories we have found that following the above advice allows continued and safe running of the Optilite lgM assay, if calibration issues are still experienced when following this advice please do not hesitate to contact a Binding Site representative. Advice on action to be taken by the user: " Follow advice
Quantity in commerce
3294 kits
Distribution
CA, PA, FL, AR, SD, MA, WV, AL, GA, IL, MN, MD, AZ, NC, TN, WA, MI, NY, KY, SC, IN, OR, NE, OH, TX, VT, WA, UT, IA,
Product codes and lots
Lot # 422251

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1327-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1327-2020Class IITerminatedVoluntary: Firm initiated