TwinFix Ti 2.8 mm HS Suture Anchor -
Smith & Nephew, Inc. began a recall of TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone on . FDA classified it as Class II and gives the reason as "Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1273-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- TwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Due to a potential for sterile barrier breach. A complaint was received that indicated the protective tube of the device came off inside the pouch enabling the pointed end of the device to puncture the package causing a breach of the sterile barrier, which could affect the sterility of the device.
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Smith Nephew "Urgent Medical Device Recall Notice R-2019-22" letter dated December 30, 2019. Customers were advised to: 1.Please inspect your inventory and locate any devices from the listed product and batch numbers on the first page of this Field Action Notification, and quarantine them immediately. a. If you are a distributor, you must notify your customers of the field action and ensure that these actions are carried out. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to [email protected] or fax to 901-566-7975. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number. If you have any questions or concerns regarding this recall please contact [email protected].
- Quantity in commerce
- 82 distributed throughout US:AZ, ID, MO, NC, ND, NE, OK, SC, TN. 243 distributed OUS: AE, AU, CA, CH, ES, FI, JP, PR, SG. Total: 325 units
- Distribution
- Worldwide distribution US Nationwide distribution in the states of AZ, ID, MO, NC, ND, NE, OK, SC, TN, and countries of AE, AU, CA, CH, ES, FI, JP, PR, SG.
- Product codes and lots
- Batch numbers: 2010547, 2025137, 2036695, 50763859, & 50773071. Part Number 72200796. UDI 10600147.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1273-2020 | Class II | Terminated | Voluntary: Firm initiated |