Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft
Cook Inc. began a recall of Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft on . FDA classified it as Class II and gives the reason as "The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1355-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Zenith Alpha Spiral-Z Endovascular Leg, Catalog Prefix ZISL. The product is an endovascular graft.
- Product code
- MIH System, Endovascular Graft, Aortic Aneurysm Treatment
- Reason for recall
- The firm has identified factors which contributed to failures relating to thrombus formation and/or lumen occlusion reports, and is highlighting to customers key points of the IFU pertaining to the prevention of these factors. In addition, the firm has updated the Planning and Sizing Worksheet to include associated information.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- On January 14, 2020, the firm notified affected customers via Urgent Medical Device Correction letters/emails. Customers were advised to pay especial attention to several aspects of the Instructions for Use (IFU) which were of key importance when implanting a device. Customers were also provided with an updated Planning and Sizing Worksheet. If you have any questions or concerns, please contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235.
- Quantity in commerce
- 68,241
- Distribution
- Foreign distribution only. Distributed worldwide.
- Product codes and lots
- All lots. OUS distribution only; Product is not cleared or distributed in the USA.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1355-2020 | Class II | Terminated | Voluntary: Firm initiated |