Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1157-2020Class II

Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD...

Implant Direct Sybron Manufacturing LLC / initiated 2019-10-11 / Terminated / root cause: packaging

Implant Direct Sybron Manufacturing LLC began a recall of Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD Platform, RxOnly, Sterile R on . FDA classified it as Class II and gives the reason as "Through a customer complaint and and internal investigation it was determined that some implant packages contain the incorrect cover screw and extender. This could result in delay in surgery and/or a potential new...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1157-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Implant Direct Sybron Manufacturing LLC
Product
Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD Platform, RxOnly, Sterile R,
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
Through a customer complaint and and internal investigation it was determined that some implant packages contain the incorrect cover screw and extender. This could result in delay in surgery and/or a potential new surgical procedures to place the correct parts. The level of severity varies based on a one or two stage surgery.
Root cause tag
packaging
FDA root cause
Mixed-up of materials/components
Action
On October 11, 2019 the firm sent an "URGENT: MEDICAL DEVICE RECALL" letter via FedEx overnight priority mail. The letter advised the consignees of the recall. In addition, the letter states: We request that you take the following actions immediately:. 1. Please review your inventory for the affected product. If located, please quarantine the affected product and return the product to us according to the instructions on the attached form. 2. Please complete and return the Acknowledgement Form within fortyeight (48) hours of receipt of this notification, whether you have affected product or not. 3. If you are an authorized Implant Direct Sybron Manufacturing LLC distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification. 4. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. Customer Care is available Monday through Friday, between the hours of 5am-5pm PST. If you or your patients have experienced any issues as a result of the affected products noted in this URGENT: MEDICAL DEVICE RECALL HHE 855710 100419 Rev 01 Status: RELEASED printed 10/10/2019 2:27:23 PM by Jose Trejo www.implantdirect.com | 888-649-6425 communication, you may voluntarily report the incident to the FDA through the MEDWATCH reporting system at the following: http://www.fda.gov/Safety/MedWatch/HowToReport/ucm053074.htm
Quantity in commerce
78 implants
Distribution
US: AR, AZ, CA, ID, IL, IN, KY, MD, ND, NH, NJ, NV, NY, Oh, OR, PA, TX, UT, WA, WI, WV, GA,MO
Product codes and lots
Lot # 68209

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1157-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1157-2020Class IITerminatedVoluntary: Firm initiated