COHERENCE Oncologist, Model Nos. 07333680 & 07351898
Siemens Medical Solutions USA, Inc began a recall of COHERENCE Oncologist, Model Nos. 07333680 & 07351898 on . FDA classified it as Class II and gives the reason as "A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1346-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- COHERENCE Oncologist, Model Nos. 07333680 & 07351898
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- A potential safety issue arises when an offset calculation is followed by a filter operation; correcting the image alignment after this sequence will result in incorrect offset values which could lead to incorrect repositioning of the patient and dose to wrong location.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- The firm started notifying customers on January 16, 2020, with letters titled Urgent Medical Device Correction. Customers were advised of the product safety issue. To avoid the possibility of an incorrect alignment of the patient, the user must ensure that filtered images arrive on the COHERENCE Oncologist, or apply any additional filters to the image BEFORE using the "interactive shift" or "landmark registration" tools. The firm will be offering a software solution for the issue free of charge. The firm's letter stated that based on their investigation, customers may continue to use their devices.
- Quantity in commerce
- 64 Worldwide/26 Domestic
- Distribution
- Domestic distribution to CA, FL, GA, IA, MA, MI, NE, NJ, NV, NY, OH, OK, PA, SD, TN, WV. Worldwide foreign distribution.
- Product codes and lots
- Software versions 2.0.49, 2.0.50, 2.0.51
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1346-2020 | Class II | Terminated | Voluntary: Firm initiated |