Z-1451-2020Class II

FilmArray Gastronintestinal (GI) Panel, REF: RFIT-ASY-0104, 6-pack kit, UDI (01)00815381020116, CE IVD, Rx...

BioFire Diagnostics, LLC / initiated 2019-08-08 / Terminated / root cause: manufacturing process

BioFire Diagnostics, LLC began a recall of FilmArray Gastronintestinal (GI) Panel, REF: RFIT-ASY-0104, 6-pack kit, UDI (01)00815381020116, CE IVD, Rx Only on . FDA classified it as Class II and gives the reason as "Complaints have been received concerning elevated rates of false negative results while using the Gastrointestinal (GI) Panel. False negative results may lead to incorrect treatment being provided which could...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1451-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
BioFire Diagnostics, LLC
Product
FilmArray Gastronintestinal (GI) Panel, REF: RFIT-ASY-0104, 6-pack kit, UDI (01)00815381020116, CE IVD, Rx Only
Product code
PCH Gastrointestinal Pathogen Panel Multiplex Nucleic Acid-Based Assay System
Reason for recall
Complaints have been received concerning elevated rates of false negative results while using the Gastrointestinal (GI) Panel. False negative results may lead to incorrect treatment being provided which could potentially lead to more serious health events.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
on August 8, 2019, a "Urgent: Medical Device Recall (Removal)" letter was sent to all affected consignees via FedEx priority-overnight. The notification informed consignees about the recall and asked they take the following actions: Requested Actions: " Please immediately examine your inventory for BioFire GI Panel reagents with kit lot number 0537319. The lot number can be found on the product label outside of the reagent box. " If reagents with the related kit lot number are found, please stop using the affected product, quarantine them, and call BioFire Technical Support to have the product returned. The BioFire Technical Support Team can assist with shipping logistics and provide a replacement. " Complete and return the enclosed Acknowledgement of Receipt Form as soon as possible and return to BioFire via one of the methods listed on the form. " If you have any questions or concerns, please dont hesitate to contact our customer support department at [email protected] or via telephone by dialing 1.800.735.6544 and selecting option 5 for product technical support.
Quantity in commerce
23 kits (690 pouches)
Distribution
US: KY, IL, NJ, MS
Product codes and lots
Kit Lot: 0537319 / Pouch Lot: 681619  Pouch Lot Date of Expiration 2020-05-30

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1451-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1451-2020Class IITerminatedVoluntary: Firm initiated