LN 3R70-01; software version 2.6.2 and earlier. The software is used with the Alinity i (LN 03R65-01)...
Abbott Gmbh & Co. KG began a recall of LN 3R70-01; software version 2.6.2 and earlier. The software is used with the Alinity i (LN 03R65-01) processing module and the Alinity c (LN 03R67-01)... on . FDA classified it as Class II and gives the reason as "Abbott has identified potential performance issues for the Alinity ci -series Software version 2.6.2 and earlier". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1580-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- LN 3R70-01; software version 2.6.2 and earlier. The software is used with the Alinity i (LN 03R65-01) processing module and the Alinity c (LN 03R67-01) processing module.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Abbott has identified potential performance issues for the Alinity ci -series Software version 2.6.2 and earlier.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated the recall letter on January 24, 2020. The "Product Correction, Urgent - Immediate Action Required" notice identified the correction of the Alinity ci-series Software version 2.6.2 and earlier due to potential performance issues. the notice informed that Abbott is releasing Alinity ci-series Software version 3.1.0 to correct the issues. The notice provided the necessary actions required to be performed until the mandatory upgrade to SW 3.1.0 is complete. The notice referenced that details are included in table1 below. the notice detail that an Abbott representative will schedule a mandatory upgrade of your Alinity ci-series to software version 3.1.0. The notice also informed the customer if you have forwarded the product listed above to other laboratories, please inform them of the Product Correction and provide them a copy of this letter. The notice further instructs the customer to please retain this letter for your laboratory records The notice instructed the customers to please Abbott's Customer Service at 1 877 4ABBOTT(available 24 hours a day, 7 days a week). For customers outside the U.S., to please contact your local area Customer Service.
- Quantity in commerce
- 2676 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, ID, IL, LA, MA, MD, ME, MI, MN, MC, MS, NC, NJ, NY, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and countries of Albania, Argentina, Australia, Austria, Bahrain, Bangledesh, Belgium, Bermuda, Bosnia, Botswana, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Madagascar, Malaysia, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago Turkey, UEA, UK, Ukraine, Uruguay, Vietnam.
- Product codes and lots
- LN 3R70-01; software version 2.6.2 and earlier.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1580-2020 | Class II | Terminated | Voluntary: Firm initiated |