Z-1499-2020Class II

HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3...

Philips North America LLC / initiated 2020-02-26 / Terminated / root cause: design

Philips North America, LLC began a recall of HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and... on . FDA classified it as Class II and gives the reason as "Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1499-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
HeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and M3536M9.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports involved a death of a patient who could not be resuscitated.
Root cause tag
design
FDA root cause
Device Design
Action
On February 26, 2020, the firm distributed Urgent Medical Device Correction letters to affected customers. Customers were informed that if the device is dropped or subjected to a severe mechanical shock, it may suffer internal damage even in the absence of visible external damage or the Ready for Use indicator unit does not immediately indicate a problem. Customers were instructed to inform all users that if the device is dropped or subjected to severe mechanical shock and the exterior case is still intact, they should immediately perform an operational check as described in the IFU section "Performing the Operational Check" in the Maintenance Chapter. The unit should be taken out of service and Philips Customer Service contacted if the unit is visibly damaged or if the device fails the operational check, i.e., if the RFU indicator changes to a red-X or the device emits a periodic audible chirp, as described in the IFU. Customers were instructed to insert a copy of the customer notice into each copy of the HeartStart MRx IFU. If you need further information or support concerning this notification, please contact your local Philips representative or call the firm at 1-800-722-9377.
Quantity in commerce
97788 units
Distribution
US Nationwide domestic distribution. Worldwide foreign distribution.
Product codes and lots
All units are affected by this recall

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1499-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1499-2020Class IITerminatedVoluntary: Firm initiated