Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C...

Abbott Medical / initiated 2020-01-22 / Terminated

St Jude Medical, Cardiac Rhythm Management Division began a recall of Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045)... on . FDA classified it as Class II and gives the reason as "Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1351-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045), CD1377-36C (05414734507622), CD2277-36Q (05414734505932), CD2377-36QC (05414734507523), with Merlin Patient Care System with software model 3330
Product code
NVZ Pulse Generator, Permanent, Implantable
Reason for recall
Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely.
FDA root cause
Other
Action
Starting on 01/22/20, Important Medical Device Correction notices were hand delivered by the recalling firm's sales representatives. Physicians were informed that the recalling firm developed a software patch for programmers which restores wireless RF communication capability in affected devices. Device explant is not required for the update. The recalling firm will assist in updating programmer software and restoring wireless RF communication for affected devices. Correction notices recommend working with firm representatives to help correct affected devices during the patient's next regularly scheduled visit. Physicians were asked to complete and return the Acknowledgement Forms and to retain the correction notice. Customers with additional questions were encouraged to call Technical Support at 1-800-722-3774.
Quantity in commerce
256
Distribution
U.S.: WV, MS, TX, DE, OH, AZ, MI, NC, KY, WA, NY, CA, IN, MO, MD, NJ, FL, PA, OK. O.U.S.: United Kingdom, Turkey, Sweden, Spain, Portugal, Poland, Philippines, Pakistan, Norway, New Zealand, Netherlands, Japan, Italy, Hong Kong, Greece, Germany, France, Finland, Denmark, Canada, Belgium, Austria, Australia
Product codes and lots
Serial #: 1138400, 1138406, 1138391, 1138393, 1138403, 1138366, 1138387, 1138396, 1138386, 1138399, 1138374, 1138362, 1138368, 1138397, 1138401, 1138372, 1138375, 1138382, 1138380, 1138405, 1138377, 1138371, 1138390, 1138398, 1138367, 1138395, 1138389, 1138379, 1138370, 1138388, 1138364, 1138373, 1138385, 1138394, 1138381, 1138404, 1138392, 1138384, 1138376, 1138369, 1138365, 1138465, 1138620, 1138573, 1138590, 1138407, 1138408, 1138409, 1138411, 1138412, 1138413, 1138415, 1138416, 1138417, 1138419, 1138421, 1138422, 1138423, 1138424, 1138429, 1138431, 1138435, 1138436, 1138437, 1138441, 1138442, 1138467, 1138468, 1138469, 1138513, 1138612, 1138613, 1138615, 1138616, 1138617, 1138624, 1138626, 1138420, 1138587, 1138628, 1138510, 1138619, 1138428, 1138432, 1138433, 1138445, 1138446, 1138447, 1138449, 1138451, 1138455, 1138466, 1138470, 1138471, 1138474, 1138479, 1138488, 1138502, 1138507, 1138523, 1138538, 1138544, 1138548, 1138556, 1138575, 1138576, 1138577, 1138580, 1138597, 1138599, 1138608, 1138495, 1138503, 1138547, 1138563, 1138565, 1138581, 1138585, 1138543, 1138461, 1138501, 1138505, 1138508, 1138545, 1138546, 1138553, 1138557, 1138566, 1138567, 1138583, 1138588, 1138486, 1138489, 1138490, 1138491, 1138496, 1138497, 1138500, 1138542, 1138549, 1138550, 1138551, 1138555, 1138559, 1138560, 1138561, 1138564, 1138570, 1138579, 1138582, 1138586, 1138611, 1138618, 1138472, 1138482, 1138595, 1138589, 1138614, 1138434, 1138440, 1138443, 1138448, 1138450, 1138452, 1138453, 1138456, 1138457, 1138458, 1138459, 1138460, 1138462, 1138463, 1138464, 1138473, 1138476, 1138478, 1138480, 1138481, 1138514, 1138515, 1138516, 1138518, 1138519, 1138520, 1138521, 1138522, 1138524, 1138525, 1138526, 1138527, 1138528, 1138529, 1138530, 1138531, 1138532, 1138533, 1138534, 1138536, 1138537, 1138539, 1138540, 1138594, 1138596, 1138600, 1138601, 1138602, 1138603, 1138604, 1138605, 1138606, 1138607, 1138609, 1138622, 1138623, 1138631, 1138632, 1138633, 1138634, 1138635, 1138636, 1138487, 1138494, 1138552, 1138554, 1138610, 1138621, 1138493, 1138499, 1138504, 1138569, 1138627, 1138484, 1138571, 1138425, 1138426, 1138438, 1138439, 1138444, 1138498, 1138492, 1138509, 1138541, 1138558, 1138562, 1138568, 1138584, 1138578, 1138511, 1138512, 1138572, 1138591, 1138592, 1138593, 1138625, 1138485, 1138574, 1138430, 1138598, 1138402, 1138363, 1138454, 1138410, 1138383, 1138378, 1138517, 1138535, 1138418, 1138506, 1138414, 1138427, 1138475, 1138477, 1138483, 1138630

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1351-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1351-2020Class IITerminatedVoluntary: Firm initiated