Z-1557-2020Class II

Optium EZ Blood Glucose Monitoring System, UDI: 00093815710653

Abbott Medical / initiated 2019-11-14 / Terminated / root cause: labeling and UDI

Abbott Diabetes Care, Inc. began a recall of Optium EZ Blood Glucose Monitoring System, UDI: 00093815710653 on . FDA classified it as Class II and gives the reason as "Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1557-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Optium EZ Blood Glucose Monitoring System, UDI: 00093815710653
Product code
NBW System, Test, Blood Glucose, Over The Counter
Reason for recall
Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
Root cause tag
labeling and UDI
FDA root cause
Labeling False and Misleading
Action
On 11/14/19, Urgent Medical Device Correction notices were mailed to customers via overnight delivery. Customers were told to continue to use instructions for use regarding the cleaning of meters for use on single patients only. Customers were further asked to complete and return customer response forms. Customers with additional questions were encouraged to call customer service in 1-888-522-5226.
Distribution
U.S.: KY, NY, TN, MA, IL, SC, VT, FL, CA, NC, MS, NJ, TX, AL, GA, OK, CT, PA, NE, ND, MT, MO, DC, ME, LA, KS, ID, OH, MN, NM, VA, IN
Product codes and lots
Not associated with specific model/lot numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1557-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1556-2020Class IITerminatedVoluntary: Firm initiated
Z-1557-2020Class IITerminatedVoluntary: Firm initiated
Z-1558-2020Class IITerminatedVoluntary: Firm initiated