FreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810
Abbott Diabetes Care, Inc. began a recall of FreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810 on . FDA classified it as Class II and gives the reason as "Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1558-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- FreeStyle Precision H Blood Glucose and Ketone Monitoring System, UDI: 00093815711810
- Product code
- JIN Nitroprusside, Ketones (Urinary, Non-Quant.)
- Reason for recall
- Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- On 11/14/19, Urgent Medical Device Correction notices were mailed to customers via overnight delivery. Customers were told to continue to use instructions for use regarding the cleaning of meters for use on single patients only. Customers were further asked to complete and return customer response forms. Customers with additional questions were encouraged to call customer service in 1-888-522-5226.
- Distribution
- U.S.: KY, NY, TN, MA, IL, SC, VT, FL, CA, NC, MS, NJ, TX, AL, GA, OK, CT, PA, NE, ND, MT, MO, DC, ME, LA, KS, ID, OH, MN, NM, VA, IN
- Product codes and lots
- All lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1556-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1557-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1558-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.