Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503
Maquet Cardiovascular Us Sales, Llc began a recall of Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to... on . FDA classified it as Class III and gives the reason as "Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1391-2020
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Getinge 46-Series Washer Disinfector Model # 46-4, Part Number:46-4-503 Product Usage: Used in facilities such as Central Sterile Supply Departments, next to the Operating Room, in dental clinics or hospital wards to clean, thermally or chemically disinfect and dry surgical goods.
- Product code
- MEC Disinfector, Medical Devices
- Reason for recall
- Due to a missing water intake for Deionized water on the Getinge 46-Series Washer Disinfector. There were no valve, hose, or clamps for this option included with the device therefore the units could not be properly installed. There have been no adverse events reported resulting in serious illness or injuries.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Getinge issued an Urgent Medical Device Recall Correction letter to U.S. consignees via FedEx Priority Overnight delivery with signature proof of delivery on January 16, 2020. Letter provides identification of the issue, health risk and action to take. Customers were advised to examine your inventory immediately, complete and sign the MEDICAL DEVICE RECALL - RESPONSE FORM to acknowledge that you have received this notification, return the completed form to Getinge Disinfection AB by e-mailing a scanned copy to [email protected] or by faxing the form to 1-800-574-5051. Should you have one or more Getinge 46-Series Washer Disinfectors (models 46-4-502 and 46-4-503) you will be contacted by a Getinge Service Representative to schedule on-site service of your device, free of charge, by a Getinge Service Representative. You can also arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9438872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone). If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action.
- Quantity in commerce
- 7 units
- Distribution
- U.S. Distribution: CA, CO, IA, MN, NY, SC, VA
- Product codes and lots
- Serial Numbers: WAA090094, WAA090112, WAA090409, WAA090521, WAA090757, WAA090805, WAA091170, WAA091288, WAA091292, WAA091454, WAA091460 WAA091463, WAA091477. UDI Code 07340153700048
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1390-2020 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-1391-2020 | Class III | Terminated | Voluntary: Firm initiated |