Z-1842-2020Class II

Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound...

Philips North America LLC / initiated 2019-08-30 / Terminated / root cause: manufacturing process

Philips Ultrasound Inc began a recall of Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging... on . FDA classified it as Class II and gives the reason as "The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1842-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies
Product code
ITX Transducer, Ultrasonic, Diagnostic
Reason for recall
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Root cause tag
manufacturing process
FDA root cause
Process control
Action
If you have an EPIQ, Affiniti or iE33 system, verify that all your S7-3t and S8-3t TEE transducers are programmed properly by connecting them to the system, checking the transducer temperature (TEE T) on the display, and following the instructions provided on this letter. If you determine that your transducer is not properly programmed, discontinue using the transducer immediately and contact your local Philips representative to schedule an appointment with a field service engineer. If you have an HD11 or CX50 system discontinue using the S7-3t and S8-3t transducer immediately and contact your local Philips representative for support. Please complete the included reply form on the last page and return to Philips as soon as possible via email to [email protected], or fax to 1-833-512-7756.
Distribution
Worldwide distribution - US Nationwide distribution.
Product codes and lots
All lots.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1842-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1841-2020Class IITerminatedVoluntary: Firm initiated
Z-1842-2020Class IITerminatedVoluntary: Firm initiated
Z-1843-2020Class IITerminatedVoluntary: Firm initiated
Z-1844-2020Class IITerminatedVoluntary: Firm initiated