Elecsys CA 19-9, Cat. No. 07027028190 -
Roche Diagnostics Operations, Inc. began a recall of Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer on . FDA classified it as Class II and gives the reason as "The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1444-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Elecsys CA 19-9, Cat. No. 07027028190 - Product Usage: the product is an immunoassay for monitoring and management of pancreatic cancer
- Product code
- NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer
- Reason for recall
- The firm has received an increased number of complaints concerning non-reproducible elevated results ("high flyers") for one lot of Elecsys CA 19-9 reagent, which may affect clinical interpretation.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On February 7, 2020, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers were informed of the product issue and given two options for CA 19-9 testing until another reagent lot is available for the cobas e 801 module: 1) Use a different Roche platform for testing, or 2) Implement a workaround, described in detail in the customer letter. The letter also provided instructions for receiving credit depending on which option was elected. Roche will notify customers when a new lot of CA 19-9 reagent for the cobas e 801 is available for ordering. ***Updated 10/5/20*** On 10/2/20, the firm issued an updated letter to customers to communicate an expansion to the recall. An additional lot was added to the recall scope.
- Quantity in commerce
- 684
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IL, IN, KS, KY, MA, ME, MI, NC, NH, NJ, NY, TN, TX, WA and WI.
- Product codes and lots
- Lot 416245 ***Updated 10/5/20*** 483123 ***Updated 8/5/22*** Additional released lots are subject to the same workaround: 46449, 50474,525510, and 568976
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1444-2020 | Class II | Terminated | Voluntary: Firm initiated |