Z-1741-2020Class II

Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power...

Siemens Healthcare Diagnostics Inc. / initiated 2019-05-17 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352 on . FDA classified it as Class II and gives the reason as "Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1741-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued letters dated 5/8/2019 on 5/17/2019 informing the customers of the issue and that a spilled sample could lead to exposure of biohazardous material and that a software update would be made by their field service engineers at the next service call.
Quantity in commerce
291 units
Distribution
Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, India, Italy, Republic of Korea, Kuwait, Malaysia, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey, UAE, United Kingdom, and Viet Nam.
Product codes and lots
All serial numbers installed before 5/20/2019:  Serials numbers in the U.S.: PW211409002, PW211501005, PW211501006, PW211501007, PW211501008, PW211501013, PW211505001, PW211508003, PW211509002, PW211510001, PW211510002, PW211511001, PW211511002, PW211511003, PW211511005, PW211512001, PW211601001, PW211601002, PW211602002, PW211602003, PW211602004, PW211603001, PW211603002, PW211603003, PW211604001, PW211604002, PW211604005, PW211604006, PW211604009, PW211604010, PW211605001, PW211605002, PW211605004, PW211605005, PW211605006, PW211605007, PW211606005, PW211606007, PW211607001, PW211607004, PW211607005, PW211607009, PW211607017, PW211607018, PW211608001, PW211608006, PW211610002, PW211610006, PW211610007, PW211610008, PW211610011, PW211611001, PW211611002, PW211611003, PW211611004, PW211611005, PW211612003, PW211612006, PW211612007, PW211612008, PW211612010, PW211612013, PW211701001, PW211701010, PW211701013, PW211703004, PW211706008, PW211706009, PW211707005, PW211707006, PW211707007, PW211707009, PW211707011, PW211707012, PW211708008, PW211708009, PW211708010, PW211709004, PW211709011, PW211710001, PW211710009, PW211710010, PW211710012, PW211711002, PW211711004, PW211711007, PW211712001, PW211802013, PW211804005, PW211804006, PW211807006, and PW211807008.  Serial numbers OUS: PW201607008, PW211409001, PW211501001, PW211501002, PW211501003, PW211501004, PW211501010, PW211501011, PW211501012, PW211506001, PW211506002, PW211506003, PW211506004, PW211506005, PW211506006, PW211506007, PW211507001, PW211507002, PW211507003, PW211507004, PW211508001, PW211508002, PW211508004, PW211509001, PW211510003, PW211511004, PW211512002, PW211512003, PW211601005, PW21160113, PW21160114, PW211602001, PW211602005, PW211602006, PW211603004, PW211603005, PW211603006, PW211604003, PW211604004, PW211604007, PW211604008, PW211605008, PW211606001, PW211606002, PW211606003, PW211606004, PW211606006, PW211606008, PW211607002, PW211607003, PW211607006, PW211607007, PW211607010, PW211607011, PW211607012, PW211607013, PW211607014, PW211607015, PW211607016, PW211608002, PW211608003, PW211608004, PW211608005, PW211610003, PW211610004, PW211610005, PW211610009, PW211610010, PW211611006, PW211612001, PW211612002, PW211612005, PW211612009, PW211612012, PW211701002, PW211701003, PW211701004, PW211701005, PW211701006, PW211701007, PW211701008, PW211701011, PW211701012, PW211701014, PW211702001, PW211702002, PW211702003, PW211702004, PW211702005, PW211702006, PW211702007, PW211702008, PW211703001, PW211703002, PW211703003, PW211703005, PW211703006, PW211703007, PW211703008, PW211703009, PW211703010, PW211703011, PW211706001, PW211706002, PW211706003, PW211706004, PW211706005, PW211706006, PW211706007, PW211706010, PW211706011, PW211706012, PW211706013, PW211707001, PW211707002, PW211707003, PW211707004, PW211707010, PW211708001, PW211708002, PW211708003, PW211708004, PW211708005, PW211708006, PW211708007, PW211709001, PW211709002, PW211709003, PW211709005, PW211709006, PW211709007, PW211709009, PW211709010, PW211709012, PW211710002, PW211710003, PW211710004, PW211710005, PW211710006, PW211710007, PW211710008, PW211710011, PW211711003, PW211712002, PW211712003, PW211712004, PW211712005, PW211712006, PW211712007, PW211712008, PW211712009, PW211712010, PW211801001, PW211801002, PW211801005, PW211801006, PW211801007, PW211801008, PW211801009, PW211801010, PW211801011, PW211801012, PW211802001, PW211802002, PW211802003, PW211802004, PW211802005, PW211802008, PW211802009, PW211802012, PW211802014, PW211802015, PW211802016, PW211803001, PW211803002, PW211803003, PW211803004, PW211803005, PW211804001, PW211804002, PW211804003, PW211804004, PW211804010, PW211804011, PW211804012, PW211804013, PW211804014, PW211804015, PW211804016, PW211804017, PW211804018, PW211804019, PW211805001, PW211805004, PW211805005, PW211805006, PW211805007, PW211806007, and PW212612004.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1741-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1741-2020Class IITerminatedVoluntary: Firm initiated