Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power...
Beckman Coulter Inc. began a recall of Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352 on . FDA classified it as Class II and gives the reason as "Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1741-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Beckman Coulter Power Express Sample Processing System AU5800XL connection unit, catalog #B36351; and Power Express Dynamic Inlet, catalog B36352.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Potential exposure to biohazard. Software design problem causes excess speed and vibrations in the unload arm movement which causes caused sample splashing.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued letters dated 5/8/2019 on 5/17/2019 informing the customers of the issue and that a spilled sample could lead to exposure of biohazardous material and that a software update would be made by their field service engineers at the next service call.
- Quantity in commerce
- 291 units
- Distribution
- Distribution was made to AL, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, MI, NC, ND, NJ, NY, OH, OR, PA, SD, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Argentina, Australia, Canada, China, Czech Republic, Egypt, France, Germany, India, Italy, Republic of Korea, Kuwait, Malaysia, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Taiwan, Turkey, UAE, United Kingdom, and Viet Nam.
- Product codes and lots
- All serial numbers installed before 5/20/2019: Serials numbers in the U.S.: PW211409002, PW211501005, PW211501006, PW211501007, PW211501008, PW211501013, PW211505001, PW211508003, PW211509002, PW211510001, PW211510002, PW211511001, PW211511002, PW211511003, PW211511005, PW211512001, PW211601001, PW211601002, PW211602002, PW211602003, PW211602004, PW211603001, PW211603002, PW211603003, PW211604001, PW211604002, PW211604005, PW211604006, PW211604009, PW211604010, PW211605001, PW211605002, PW211605004, PW211605005, PW211605006, PW211605007, PW211606005, PW211606007, PW211607001, PW211607004, PW211607005, PW211607009, PW211607017, PW211607018, PW211608001, PW211608006, PW211610002, PW211610006, PW211610007, PW211610008, PW211610011, PW211611001, PW211611002, PW211611003, PW211611004, PW211611005, PW211612003, PW211612006, PW211612007, PW211612008, PW211612010, PW211612013, PW211701001, PW211701010, PW211701013, PW211703004, PW211706008, PW211706009, PW211707005, PW211707006, PW211707007, PW211707009, PW211707011, PW211707012, PW211708008, PW211708009, PW211708010, PW211709004, PW211709011, PW211710001, PW211710009, PW211710010, PW211710012, PW211711002, PW211711004, PW211711007, PW211712001, PW211802013, PW211804005, PW211804006, PW211807006, and PW211807008. Serial numbers OUS: PW201607008, PW211409001, PW211501001, PW211501002, PW211501003, PW211501004, PW211501010, PW211501011, PW211501012, PW211506001, PW211506002, PW211506003, PW211506004, PW211506005, PW211506006, PW211506007, PW211507001, PW211507002, PW211507003, PW211507004, PW211508001, PW211508002, PW211508004, PW211509001, PW211510003, PW211511004, PW211512002, PW211512003, PW211601005, PW21160113, PW21160114, PW211602001, PW211602005, PW211602006, PW211603004, PW211603005, PW211603006, PW211604003, PW211604004, PW211604007, PW211604008, PW211605008, PW211606001, PW211606002, PW211606003, PW211606004, PW211606006, PW211606008, PW211607002, PW211607003, PW211607006, PW211607007, PW211607010, PW211607011, PW211607012, PW211607013, PW211607014, PW211607015, PW211607016, PW211608002, PW211608003, PW211608004, PW211608005, PW211610003, PW211610004, PW211610005, PW211610009, PW211610010, PW211611006, PW211612001, PW211612002, PW211612005, PW211612009, PW211612012, PW211701002, PW211701003, PW211701004, PW211701005, PW211701006, PW211701007, PW211701008, PW211701011, PW211701012, PW211701014, PW211702001, PW211702002, PW211702003, PW211702004, PW211702005, PW211702006, PW211702007, PW211702008, PW211703001, PW211703002, PW211703003, PW211703005, PW211703006, PW211703007, PW211703008, PW211703009, PW211703010, PW211703011, PW211706001, PW211706002, PW211706003, PW211706004, PW211706005, PW211706006, PW211706007, PW211706010, PW211706011, PW211706012, PW211706013, PW211707001, PW211707002, PW211707003, PW211707004, PW211707010, PW211708001, PW211708002, PW211708003, PW211708004, PW211708005, PW211708006, PW211708007, PW211709001, PW211709002, PW211709003, PW211709005, PW211709006, PW211709007, PW211709009, PW211709010, PW211709012, PW211710002, PW211710003, PW211710004, PW211710005, PW211710006, PW211710007, PW211710008, PW211710011, PW211711003, PW211712002, PW211712003, PW211712004, PW211712005, PW211712006, PW211712007, PW211712008, PW211712009, PW211712010, PW211801001, PW211801002, PW211801005, PW211801006, PW211801007, PW211801008, PW211801009, PW211801010, PW211801011, PW211801012, PW211802001, PW211802002, PW211802003, PW211802004, PW211802005, PW211802008, PW211802009, PW211802012, PW211802014, PW211802015, PW211802016, PW211803001, PW211803002, PW211803003, PW211803004, PW211803005, PW211804001, PW211804002, PW211804003, PW211804004, PW211804010, PW211804011, PW211804012, PW211804013, PW211804014, PW211804015, PW211804016, PW211804017, PW211804018, PW211804019, PW211805001, PW211805004, PW211805005, PW211805006, PW211805007, PW211806007, and PW212612004.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1741-2020 | Class II | Terminated | Voluntary: Firm initiated |