Bard Biopsy EnCor Probe, MRI, 7G Trocar Tip, REF number ECPMR017G, packaged individually in sterile pouches...
Bard Peripheral Vascular Inc began a recall of Bard Biopsy EnCor Probe, MRI, 7G Trocar Tip, REF number ECPMR017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile on . FDA classified it as Class II and gives the reason as "Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1432-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- Bard Biopsy EnCor Probe, MRI, 7G Trocar Tip, REF number ECPMR017G, packaged individually in sterile pouches, 5 pouches/carton, Single Use, Rx, Sterile.
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- Lot numbers and products inadvertently not included in the scope of the previous recall. Original recall was conducted due to an increase in probe failures for leaks, suction issues, and failure to obtain samples.
- FDA root cause
- Component change control
- Action
- The recalling firm issued letters dated 1/3/2020 to each of the U.S. consignees via FedEx with proof of delivery notification. The letter identified the affected product, problem and actions to be taken. For questions call 1-800-321-4254 Option #5.
- Quantity in commerce
- 1,585 units
- Distribution
- Distribution was made to AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, OR, PA, SC, TX, VA, and WV. There was government distribution and no military distribution. Foreign distribution was made to Canada, Australia, Austria, Belgium, China, Czech Republic, Germany, France, Georgia, Greece, Hong Kong, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, and United Kingdom.
- Product codes and lots
- Lot numbers VTCT0272 and VTCV0384. UDI numbers (01)00801741086328(17)200601(10)VTCT0272 and (01)00801741086328(17)200801(10)VTCV0384.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1431-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1432-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1433-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1434-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.