Z-0283-2021Class II

da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is...

Intuitive Surgical, Inc. / initiated 2019-06-17 / Terminated / root cause: manufacturing process

Intuitive Surgical, Inc. began a recall of da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number... on . FDA classified it as Class II and gives the reason as "One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0283-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci Surgical System, IS 4000; Endoscope Controller, Model Number: 372601-16 (UDI: 00886874114193) is included in the IS 4000 Vision Cart, Model Number: 381121-31 (UDI 00886874110898) - Product Usage: used with an endoscope, surgical instruments, and accessories to perform minimally invasive surgical procedures.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Send formal customer letter communication to the affected site. The firm reported that a representative scheduled an on-site visit to replace the affected unit.
Quantity in commerce
1 system
Distribution
US Nationwide distribution including in the state of Washington.
Product codes and lots
IS4000 Endoscope Controller, Serial Number: 432242 installed on the da Vinci Surgical System, Serial Number: SK2137

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0283-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0283-2021Class IITerminatedVoluntary: Firm initiated