Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
Philips North America, LLC began a recall of Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044) on . FDA classified it as Class II and gives the reason as "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1487-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.
- Quantity in commerce
- 70
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Product codes and lots
- Serial Codes 3217017 200 58 72 CV.170 180 524615 47 211 364 30 3862570/0011 529 S.N. 519 S01H060871 505/0050114 S01H057274 01H077213 530 532 S01HA52566 348 476 320 481 467 284 352 359 459 73*004 733 536 CV.474 454 49 281 51 186 122 250 37 22 CV.312 296 373 CV.596 176 30 580 279
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1486-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1487-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1488-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1489-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1490-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1491-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1492-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1493-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1494-2020 | Class II | Terminated | Voluntary: Firm initiated |