Z-1487-2020Class II

Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)

Philips North America LLC / initiated 2019-12-30 / Terminated / root cause: design

Philips North America, LLC began a recall of Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044) on . FDA classified it as Class II and gives the reason as "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1487-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Integris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Root cause tag
design
FDA root cause
Component design/selection
Action
On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.
Quantity in commerce
70
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Product codes and lots
Serial Codes  3217017 200 58 72 CV.170 180 524615 47 211 364 30 3862570/0011 529 S.N. 519 S01H060871 505/0050114 S01H057274 01H077213 530 532 S01HA52566 348 476 320 481 467 284 352 359 459 73*004 733 536 CV.474 454 49 281 51 186 122 250 37 22 CV.312 296 373 CV.596 176 30 580 279

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1487-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1486-2020Class IITerminatedVoluntary: Firm initiated
Z-1487-2020Class IITerminatedVoluntary: Firm initiated
Z-1488-2020Class IITerminatedVoluntary: Firm initiated
Z-1489-2020Class IITerminatedVoluntary: Firm initiated
Z-1490-2020Class IITerminatedVoluntary: Firm initiated
Z-1491-2020Class IITerminatedVoluntary: Firm initiated
Z-1492-2020Class IITerminatedVoluntary: Firm initiated
Z-1493-2020Class IITerminatedVoluntary: Firm initiated
Z-1494-2020Class IITerminatedVoluntary: Firm initiated