Z-1488-2020Class II

Integris Allura 9, system codes 722018 722021

Philips North America LLC / initiated 2019-12-30 / Terminated / root cause: design

Philips North America, LLC began a recall of Integris Allura 9, system codes 722018 722021 on . FDA classified it as Class II and gives the reason as "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1488-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Integris Allura 9, system codes 722018 722021
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Root cause tag
design
FDA root cause
Component design/selection
Action
On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.
Quantity in commerce
20
Distribution
Nationwide domestic distribution. Foreign distribution worldwide.
Product codes and lots
Serial Codes  8 114 64 906187 41 38 119  75 25   87  67  107  32 97

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1488-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1486-2020Class IITerminatedVoluntary: Firm initiated
Z-1487-2020Class IITerminatedVoluntary: Firm initiated
Z-1488-2020Class IITerminatedVoluntary: Firm initiated
Z-1489-2020Class IITerminatedVoluntary: Firm initiated
Z-1490-2020Class IITerminatedVoluntary: Firm initiated
Z-1491-2020Class IITerminatedVoluntary: Firm initiated
Z-1492-2020Class IITerminatedVoluntary: Firm initiated
Z-1493-2020Class IITerminatedVoluntary: Firm initiated
Z-1494-2020Class IITerminatedVoluntary: Firm initiated