Z-1635-2020Class II

8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI...

Trilliant Surgical LLC / initiated 2019-11-04 / Terminated / root cause: sterility

Trilliant Surgical, LLC began a recall of 8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202 on . FDA classified it as Class II and gives the reason as "The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1635-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Trilliant Surgical LLC
Product
8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202
Product code
HTY Pin, Fixation, Smooth
Reason for recall
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Root cause tag
sterility
FDA root cause
Mixed-up of materials/components
Action
The recall strategy included sending complete recall communication and response forms as of 11/04/2019 via FedEx (Express Standard Overnight) to all consignees that showed possession / had been sold a device as of 10/30/2019. The recall communication included the following actions to be taken by the distributor/customer: 1. Check inventory stock for the identified device. 2. Open the shipper box* on the end opposite of the label and remove the internal packaging. *Note: Opening the shipper box does not disrupt the sterile barrier. The double sterile barrier is maintained within the internal packaging. 3. Through the transparent internal packaging, verify that the kit contains the appropriate quantity and type of each reamer; (1) Concave Reamer and (1) Convex Reamer. If product in possession IS identified as affected by containing (2) of the same reamer type, quarantine and/or discontinue use of the device and return the affected device(s) to the firm in accordance with the included Return Response (Acknowledgment and Receipt Form). If product in possession IS NOT affected (i.e. there is 1 Concave Reamer and 1 Convex Reamer within the packaging, complete and return the included Return Response (Acknowledgment and Receipt Form). THE INCLUDED RETURN RESPONSE IS REQUIRED EVEN IF AFFECTED PRODUCT IS NOT IDENTIFIED. For questions regarding the communication, the phone number is 800.495.2919.
Quantity in commerce
117
Distribution
Distributors in 21 states: AR, AZ, CA, FL, GA, IA,IN, KS, KY, MN, MO, NC, ND, NJ, NY, OH, TN, TX, UT, WA, and WI.
Product codes and lots
Part Number: 220-08-003, Lot Number: TSL007202;Expiration Date: 12/25/2023.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1635-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1635-2020Class IITerminatedVoluntary: Firm initiated