Z-1490-2020Class II
OmniDiagnost Eleva System codes 708028 708027
Philips North America LLC / initiated 2019-12-30 / Terminated / root cause: design
Philips North America, LLC began a recall of OmniDiagnost Eleva System codes 708028 708027 on . FDA classified it as Class II and gives the reason as "A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1490-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- OmniDiagnost Eleva System codes 708028 708027
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On December 30, 2019, the firm notified customers via Medical Device Correction letter. Customers were informed of the product issue and advised to do the following: If the capacitor of the Velara X-Ray fails as described (capacitor failure after large number of surges in a short period of time), please switch off the system, take it out of service immediately, and contact your local Philips representative. The firm will be resolving the issue by replacing a Printed Circuit Board (PCB) in the convertor, which prevents the capacitor from failing. PCB replacements started in January 2020. Customers will be contacted by their local Philips representatives to schedule a date for the PCB replacement. Customers with further questions should contact the Technical Support Line: 1-800-722-9377.
- Quantity in commerce
- 69
- Distribution
- Nationwide domestic distribution. Foreign distribution worldwide.
- Product codes and lots
- Serial codes 152 49 514 179 611 C@23273 46 560 555 464 14 90 198 15 326 143*003 451 267 FR90-111 417 59 390 164 135 23 143 80 219 24 78 73 15 79 109 624 71 44 576 70 460 62 98 63 620 147 100 68 122 422 12 54 490 97 138 20 130 67 ON100031011SN273 241 435 125 89 47 17 33 134 574
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1486-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1487-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1488-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1489-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1490-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1491-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1492-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1493-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1494-2020 | Class II | Terminated | Voluntary: Firm initiated |