VITEK 2 card AST-N371 -
bioMerieux, Inc. began a recall of VITEK 2 card AST-N371 - Product Usage: The VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the... on . FDA classified it as Class II and gives the reason as "The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1534-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- bioMerieux, Inc.
- Product
- VITEK 2 card AST-N371 - Product Usage: The VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
- Product code
- N/A
- Reason for recall
- The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- bioMerieux notified customers on about 02/05/2020 via "Urgent Product Correction Notice." The letter requested customers to check their inventory, destroy any recalled product, ensure the recall notice has been distributed to all appropriate personnel within your organization, and complete and return the attached Acknowledgement form.
- Quantity in commerce
- 4,130 kits (82,600 cards)
- Distribution
- International distribution in the countries of Germany and the Netherlands.
- Product codes and lots
- REF: 422024 Lot Number: 0210932204
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1534-2020 | Class II | Terminated | Voluntary: Firm initiated |