Z-1554-2020Class II
NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety)...
Hill-Rom Inc. / initiated 2020-02-14 / Terminated / root cause: software
Hill-Rom Inc. began a recall of NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1... on . FDA classified it as Class II and gives the reason as "There is a software issue which may result in failure to monitor or control patient's bed exit mechanism". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1554-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1. VersaCare Revision A-J with upgrade kit Revision K with UCB 1.18 2. Centrella (all versions) 3. Progressa (firmware version 1.15.0.0)
- Product code
- KMI Monitor, Bed Patient
- Reason for recall
- There is a software issue which may result in failure to monitor or control patient's bed exit mechanism.
- Root cause tag
- software
- FDA root cause
- Software Manufacturing/Software Deployment
- Action
- The firm issued a Urgent Medical Device notification on 02/14/2020 by letter. The letter identified the reason and requested the following actions be taken: 1. Hillrom is recommending customers disable the NPS software module, including monitoring for Bed Exit Not Armed, to avoid the potential for staff being unaware of the bed exit state. 2. With the NPS software disabled, caregivers will need to return to manually arming the bed exit to alert for patients exiting the bed. Hillrom is developing a software update for NaviCare Patient Safety and anticipates availability in April 2020. Next Steps: Hillrom will be reaching out to assist you with disabling the software module and plan your upgrade. If you have any questions, or would like to proactively schedule an upgrade, please contact Hillrom Technical Support at 1-800-445-3720, option 3 (Cary), option 1 (NaviCare Nurse Call).
- Quantity in commerce
- 35 Software Systems
- Distribution
- Colorado, Georgia, Indiana, Louisiana, North Carolina, Nevada, Oregon, South Carolina, Tennessee, Virginia, Washington.
- Product codes and lots
- NaviCare Patient Safety Software versions 3.5.400 to 3.9.300