Z-1554-2020Class II

NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety)...

Hill-Rom Inc. / initiated 2020-02-14 / Terminated / root cause: software

Hill-Rom Inc. began a recall of NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1... on . FDA classified it as Class II and gives the reason as "There is a software issue which may result in failure to monitor or control patient's bed exit mechanism". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1554-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
NaviCare Patient Safety Software Versions 3.5.400-3.9.300. The NaviCare Patient Safety (Patient Safety) Software is compatible with the following beds: 1. VersaCare Revision A-J with upgrade kit Revision K with UCB 1.18 2. Centrella (all versions) 3. Progressa (firmware version 1.15.0.0)
Product code
KMI Monitor, Bed Patient
Reason for recall
There is a software issue which may result in failure to monitor or control patient's bed exit mechanism.
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
The firm issued a Urgent Medical Device notification on 02/14/2020 by letter. The letter identified the reason and requested the following actions be taken: 1. Hillrom is recommending customers disable the NPS software module, including monitoring for Bed Exit Not Armed, to avoid the potential for staff being unaware of the bed exit state. 2. With the NPS software disabled, caregivers will need to return to manually arming the bed exit to alert for patients exiting the bed. Hillrom is developing a software update for NaviCare Patient Safety and anticipates availability in April 2020. Next Steps: Hillrom will be reaching out to assist you with disabling the software module and plan your upgrade. If you have any questions, or would like to proactively schedule an upgrade, please contact Hillrom Technical Support at 1-800-445-3720, option 3 (Cary), option 1 (NaviCare Nurse Call).
Quantity in commerce
35 Software Systems
Distribution
Colorado, Georgia, Indiana, Louisiana, North Carolina, Nevada, Oregon, South Carolina, Tennessee, Virginia, Washington.
Product codes and lots
NaviCare Patient Safety Software versions 3.5.400 to 3.9.300

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1554-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1554-2020Class IITerminatedVoluntary: Firm initiated