Z-1547-2020Class II

Chemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 -

Randox Laboratories, Limited / initiated 2020-02-18 / Open, Classified

Randox Laboratories Ltd. began a recall of Chemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 - Product Usage: This product is intended for in vitro diagnostic use, in the quality control... on . FDA classified it as Class II and gives the reason as "The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1547-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Randox Laboratories, Limited
Product
Chemistry QC Premium Plus, Cat. No. LAL4213, GTIN 05055273209006 - Product Usage: This product is intended for in vitro diagnostic use, in the quality control of diagnostic assays. The device is for the control of accuracy.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
The control target and range value for Sodium using the ISE indirect method has been assigned incorrectly.
FDA root cause
Under Investigation by firm
Action
On February 18, 2020, the firm notified customers of the action via emailed Urgent Medical Device Correction letters. Customers were informed of the incorrectly assigned control target and range value for Sodium using the ISE indirect method. The letter stated that the sections for Roche Cobas series and mean of all instruments has now been updated, and the updated value sheets were included with the email. The updated value sheets may also be found on www.randox.com. Customers were asked to take the following actions: - Review results generated with the affected batches in line with the clinical profile of the patient. -Discuss the contents of the notice with your Medical Director. -Complete and return the included response form within 5 working days. Distributors were asked to forward a copy of the notice to all affected customers. If you have questions or concerns, please contact Randox Technical Services: 866-4-RANDOX, or [email protected]
Quantity in commerce
186 total, 27 USA
Distribution
Worldwide distribution - US Nationwide distribution in the states of WV, IN, VA, ID, PR, DE, and countries of West Indies, Czech Republic, France, Germany, Hong Kong, Ireland, Italy, Norway, Philippines, South Africa, South Korea, Sweden, Taiwan (R.O.C.), Thailand, UAE, UK.
Product codes and lots
Lot 306UL

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1547-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1547-2020Class IIOngoingVoluntary: Firm initiated