Z-1535-2020Class II

iChemVELOCITY Strips, Catalog 800-7212

Siemens Healthcare Diagnostics Inc. / initiated 2020-02-17 / Terminated

Beckman Coulter, Inc. began a recall of iChemVELOCITY Strips, Catalog 800-7212 on . FDA classified it as Class II and gives the reason as "The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1535-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
iChemVELOCITY Strips, Catalog 800-7212
Product code
KQO Automated Urinalysis System
Reason for recall
The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
FDA root cause
Under Investigation by firm
Action
An Urgent Medical Device Recall letter will be sent during the week of 02/17/2020 to the affected customers via email and mail. Customers are advised to discard the affected product and order a replacement from BEC.
Quantity in commerce
10822 vials
Distribution
Distribution in US including Puerto Rico, Canada, and Taiwan
Product codes and lots
Catalog 800-7212, UDI 10837461002628; Lot 7212269M

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1535-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1535-2020Class IITerminatedVoluntary: Firm initiated