Z-1511-2020Class II

Boston Scientific, LATITUDE Programming System, Model 3300 -

Boston Scientific Corporation / initiated 2020-02-12 / Terminated / root cause: software

Boston Scientific Corporation began a recall of Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital... on . FDA classified it as Class II and gives the reason as "Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1511-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific, LATITUDE Programming System, Model 3300 - Product Usage:The Boston Scientific LATITUDE Programming System is intended for use in hospital and clinical environments to communicate with Boston Scientific implantable systems. The software in use controls all communication functions for the Pulse Generator (PG). Detailed software applications instructions can be found in the associated product literature for the PG being interrogated.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
The firm initiated the recall by letter on 12/12/2020. The letter explained the error and requested the consignee segregate the units for removal. The firm will provide another model of programmer to be used in the interim.
Quantity in commerce
4 devices
Distribution
US Nationwide distribution in the states of NC, TN, and MO.
Product codes and lots
Serial Numbers: 008885, 010823, 010615, 007886

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1511-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1511-2020Class IITerminatedVoluntary: Firm initiated
Z-1512-2020Class IITerminatedVoluntary: Firm initiated