Z-1544-2020Class II
Medline Stick Sponge Presaturated W/PVP REF DYND70288 DYND70288H 2/PK Expiration 06/30/2022
MEDLINE INDUSTRIES, LP - Northfield / initiated 2020-02-10 / Terminated / root cause: packaging
Medline Industries Inc began a recall of Medline Stick Sponge Presaturated W/PVP REF DYND70288 DYND70288H 2/PK Expiration 06/30/2022 on . FDA classified it as Class II and gives the reason as "Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1544-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Stick Sponge Presaturated W/PVP REF DYND70288 DYND70288H 2/PK Expiration 06/30/2022.
- Product code
- OJU Skin Prep Tray
- Reason for recall
- Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Medline notified customers on February 18, 2020 via "Urgent Recall Immediate Action required" letter. The consignees were instructed to complete the response form and return the product. The recall letter identified the affected product and lot numbers. Reason for the correction was also noted in the recall letter.
- Quantity in commerce
- 1,541,532 units
- Distribution
- US Nationwide: MA, LA, IL, MO, MI, MN, NJ, KS, GA, NY, PA, SD, CT, TN, MD, OH, VA, FL, TX, KY, CO, IN, AL, CA, MT, OR, AK, AL, DC, IA, ME, NV, NE, SC, WV, WY, RI, NC, AZ, UT, DE and WI.
- Product codes and lots
- REF DYND70288 2/PK Expiration 06/30/2022
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1544-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1545-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1546-2020 | Class II | Terminated | Voluntary: Firm initiated |