Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1602-2020Class II

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000

STRYKER Communications, Inc. / initiated 2019-10-11 / Terminated

Stryker Communications began a recall of Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000 on . FDA classified it as Class II and gives the reason as "The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1602-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STRYKER Communications, Inc.
Product
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.
FDA root cause
Other
Action
On October 11, 2019, a "Urgent Medical Device Correction" letter was sent to all affected customers via FedEx. In addition to explaining the recall, the recall letter as customers to take the following action: Please assure this notification letter is disseminated within your organization. A Stryker Representative will begin scheduling visits to customer facilities to perform inspections. This correction is expected to last one year. During the inspection, the Stryker Representative will perform a visual inspection on the lower portion of the Chromophare Ceiling Mounted Surgical Light Suspension. If a nonconforming unit is found, Stryker will replace the non-conforming Chromophare Ceiling Mounted Surgical Light Suspension with a new Chromophare Ceiling Mounted Surgical Light Suspension. In the meantime, the lights can continue to remain in full use. If you have any questions or notice any issues regarding this correction: " Stryker Technical Support at 866-841-5663 or [email protected] " Nicole Kubat, Senior RA Compliance Specialist at 469-470-4318 or [email protected] " E-mail: [email protected]
Quantity in commerce
1415 Surgical Lights
Distribution
US: Nationwide OUS: Canada
Product codes and lots
Serial Number Range: 091316 70016 through 083117-73252

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1602-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1602-2020Class IITerminatedVoluntary: Firm initiated