Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod...
Medtronic Inc. began a recall of Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices on . FDA classified it as Class II and gives the reason as "Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2106-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.
- Product code
- MDS Sensor, Glucose, Invasive
- Reason for recall
- Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Urgent Medical Device Safety Notification Letters were sent to patients, health care providers (HCPs), distributors and clinical via email beginning on May 20, 2019 and via USPS certified mail beginning on May 22, 2019 notifying them of a potential issue with the transmitter battery life of the Connect Continuous Glucose Monitoring System. Customers using an Apple device with iOS version 12 or higher were advised to update to the latest version (3.2.4) of the Guardian Connect app on their iPhone as soon as possible. Customers not using iOS 12 or higher were encouraged to update to the latest version of the Guardian Connect app. Customers receiving emails were requested to submit email acknowledgements indicating they have read and understood the instructions provided in the Urgent Safety Notice Letter clicking on the button as evidence of receipt. Customers receiving paper-copy notifications were instructed to fill out a response card indicating they had read and understood the notice, and mail the card back with a pre-paid envelope. Customers with invalid email and/or physical addresses will be contacted by telephone. Customers with questions were encouraged to contact 24-Hour Technical Support Team at 1-800-646-4633.
- Quantity in commerce
- 23,978
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model Number CSS7200 (Guardian Connect App: iOS); Catalog Numbers: MMT-7821 (Guardian Connect Transmitter - GST4C); UPN - 00763000186333. There are no assigned lot or serial numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2106-2020 | Class II | Ongoing | Voluntary: Firm initiated |