Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1603-2020Class II

Avalon Elite Vascular Access Kit, Model No. 70106.3540, UDI 04037691831039

Maquet Cardiovascular Us Sales, LLC / initiated 2020-02-20 / Terminated

Maquet Cardiovascular Us Sales, Llc began a recall of Avalon Elite Vascular Access Kit, Model No. 70106.3540, UDI 04037691831039 on . FDA classified it as Class II and gives the reason as "A lot of Avalon Elite Vascular Access Kits (VAK) are being recalled due to a material mix-up with the needle portion of the kit; 81 of 255 kits were assembled with an incorrect Arterial Access Needle instead of a...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1603-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Avalon Elite Vascular Access Kit, Model No. 70106.3540, UDI 04037691831039
Product code
DRE Dilator, Vessel, For Percutaneous Catheterization
Reason for recall
A lot of Avalon Elite Vascular Access Kits (VAK) are being recalled due to a material mix-up with the needle portion of the kit; 81 of 255 kits were assembled with an incorrect Arterial Access Needle instead of a Multi-Purpose Procedure Needle.
FDA root cause
Under Investigation by firm
Action
On February 20, 2020, the recalling firm notified affected customers via Urgent Medical Device Recall letters sent through FedEx. Customers were informed of the material mix-up with the needle portion of the kit. Customers were instructed to immediately examine their inventory to determine if they have any of the affected lot. If so, please remove the kits from areas of use and return the product to Getinge following the instructions in the letter. Unused and unexpired affected product that is returned to Getinge is eligible for full credit.
Quantity in commerce
255
Distribution
Domestic distribution only to the following states: AL AZ CA CT FL GA IL MA MI NC NH NM NY OH OK TN TX VA
Product codes and lots
Lot 19-03466

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1603-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1603-2020Class IITerminatedVoluntary: Firm initiated