Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE...
Stryker Corporation began a recall of Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was... on . FDA classified it as Class II and gives the reason as "Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1509-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- On September 25, 2019 and November 26, 2019, the firm sent an "Urgent Medical Device Recall Notification" letter to all consignees via USPS mail. In addition, to informing consignees about the recall, they recall letter asked consignees to take the following actions: Actions to be taken by the Customer/User: 1. Inform individuals within your organization who need to be aware of this device removal. 2. Check all stock areas and/or operating room storage to determine if any devices with the affected Strykeprobe Sheath lot numbers from Attachment A are at your facility. - a. If you would like a list of all affected probe lot numbers, please email [email protected] 3. If affected product is found, segregate the product and call Stryker customer service at 1-800-624-4422 (Option 3) or email [email protected] to arrange for product return and issuance of credit or replacement (upon availability). - a. Remove the affected sheath, quarantine, and discontinue use of the identified non-conforming recalled sheath(s). - b. When returning units, please enclose Business Reply Form on Attachment B. 4. If affected product is NOT found: - a. Please complete Business Reply Form on Attachment B and return to [email protected] 5. Please forward a copy of this letter to any other personnel within your facility you deem appropriate. 6. Please send any questions to [email protected].
- Distribution
- OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea
- Product codes and lots
- Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1508-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1509-2020 | Class II | Terminated | Voluntary: Firm initiated |