Z-1509-2020Class II

Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE...

Stryker Corp. / initiated 2019-09-12 / Terminated / root cause: packaging

Stryker Corporation began a recall of Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was... on . FDA classified it as Class II and gives the reason as "Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1509-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker, REF: 0250070450, 5.0 mm x 45 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051926 This products was distributed with the following sheath/probe: 0250070451-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Spatula-Tip, 0250070452-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, J-TIP, 0250070453-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, L-TIP, 0250070455-PKG, 5MM X 45CM StrykeProbe Electrosurgical Probe, Needle Tip
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
Root cause tag
packaging
FDA root cause
Other
Action
On September 25, 2019 and November 26, 2019, the firm sent an "Urgent Medical Device Recall Notification" letter to all consignees via USPS mail. In addition, to informing consignees about the recall, they recall letter asked consignees to take the following actions: Actions to be taken by the Customer/User: 1. Inform individuals within your organization who need to be aware of this device removal. 2. Check all stock areas and/or operating room storage to determine if any devices with the affected Strykeprobe Sheath lot numbers from Attachment A are at your facility. - a. If you would like a list of all affected probe lot numbers, please email [email protected] 3. If affected product is found, segregate the product and call Stryker customer service at 1-800-624-4422 (Option 3) or email [email protected] to arrange for product return and issuance of credit or replacement (upon availability). - a. Remove the affected sheath, quarantine, and discontinue use of the identified non-conforming recalled sheath(s). - b. When returning units, please enclose Business Reply Form on Attachment B. 4. If affected product is NOT found: - a. Please complete Business Reply Form on Attachment B and return to [email protected] 5. Please forward a copy of this letter to any other personnel within your facility you deem appropriate. 6. Please send any questions to [email protected].
Distribution
OUS: Netherlands, Spain, Switzerland, Colombia, Canada, Japan, Taiwan,, Puerto Rico, South Africa, and South Korea
Product codes and lots
Lot Numbers: 010818-02, 010818-05, 010818-06, 010818-07, 010818-08, 010818-09, 040816-01, 040816-02

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1509-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1508-2020Class IITerminatedVoluntary: Firm initiated
Z-1509-2020Class IITerminatedVoluntary: Firm initiated