NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument...
Smith & Nephew, Inc. began a recall of NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument designed to, when used as instructed, provide the... on . FDA classified it as Class II and gives the reason as "Potential for the NAVIO Soft Tissue Protector to become stuck or bound to the bone pin intraoperatively when inserting the NAVIO bone pins". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1634-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- NAVIO Soft Tissue Protector, included in instrument kits. The NAVIO Tissue Protector is a reusable instrument designed to, when used as instructed, provide the required bone pin spacing for the Navio bone clamps and potentially reduce the amount of patient tissue affected when placing the bone pins.
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- Potential for the NAVIO Soft Tissue Protector to become stuck or bound to the bone pin intraoperatively when inserting the NAVIO bone pins.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Smith & Nephew, notified customers on about 03/03/2020 via "Urgent Medical Device Recall Notice" letter. The customers were Instructed to examine inventory and quarantine affected devices, notify customers if further distributed, and complete and return the Response Form to [email protected] or fax to 1-901-566-7975. Questions and concerns can be directed to EieldActions@smlth-nephew .com. or 901-399-1970.
- Quantity in commerce
- 348 units
- Distribution
- Nationwide distribution to AL, AZ, CA, CT, FL, GA, HI, IA, IL, KS, LA, MA, MI, MN, MS, MT, NC, NJ, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV. International distribution to Turkey, United Kingdom, Thailand, Switzerland, France, Australia.
- Product codes and lots
- Component Product Number: PFSR101092; a. Instrument Kit Number: PFSR02050, Lot Numbers 2067, 2212, 3809, 3869, 4354, 5228, 5337, 5494, 5514, 5887, 6248; b. Instrument Kit Number: PFSR02051, Lot Numbers 4354, 5514; c. Instrument Kit Number: PFSR02052, Lot Numbers 4354, 5494, 6248; d. Instrument Kit Number: PFSR02053, Lot Numbers 6248
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1634-2020 | Class II | Terminated | Voluntary: Firm initiated |