OsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number 1129957) -
Skeletal Kinetics, Llc began a recall of OsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number 1129957) - Product Usage: The product provides a bone void filler... on . FDA classified it as Class II and gives the reason as "A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0866-2021
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Skeletal Kinetics, LLC
- Product
- OsteoVation, Inject Bone Filler Kits, 5cc (part number 390-2005, manufacturing lot number 1129957) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.
- Product code
- GXP Methyl Methacrylate For Cranioplasty
- Reason for recall
- A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On June 7, 2019, customers were notified by letter sent by FedEx that Skeletal Kinetics LLC (Skeletal Kinetics) is voluntarily recalling certain OsteoVation, Inject Bone Void Filler kits distributed by OsteoMed LLC.The recall was initiated to correct a discrepancy between the Mixing Instruction Card and the Instructions for Use. The letter notes that Skeletal Kinetics has not received any customer complaints or reports of injury to the patient or end user. Customers are requested to: (1) review this letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or the enclosed pre-paid envelope; and (2) return unused inventory via Federal Express with reference to Return Material Authorization FER 3003890476-6/6/19-001-R or by pickup by the sales representative indicated on the Acknowledgement and Receipt Form. Customers with questions are instructed to contact Mr. Randy Mosley, Director of QA, Skeletal Kinetics LLC by telephone at 972 677 4789 Monday-Friday, 8 AM 5 PM CDT.
- Quantity in commerce
- 2 units
- Distribution
- U.S. Nationwide distribution including in the states of TX, NY, FL and AZ.
- Product codes and lots
- OsteoVation, Inject, 5cc (part number 390-2005, manufacturing lot number 1129957, label no. 12464) . UDI: (01) 0 0813845 02005 4 (17) 210401 (10) 1129957
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0866-2021 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0867-2021 | Class II | Terminated | Voluntary: Firm initiated |