Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0867-2021Class II

OsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) -

Skeletal Kinetics, LLC / initiated 2019-06-06 / Terminated / root cause: labeling and UDI

Skeletal Kinetics, Llc began a recall of OsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) - Product Usage: The product provides a bone void filler that is... on . FDA classified it as Class II and gives the reason as "A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0867-2021
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Skeletal Kinetics, LLC
Product
OsteoVation, Inject, Bone Filler Kits 10cc (part number 390-2010, manufacturing lot 1130303) - Product Usage: The product provides a bone void filler that is replaced by bone during the healing process.
Product code
GXP Methyl Methacrylate For Cranioplasty
Reason for recall
A discrepancy was found between the Mixing Instruction Card and the Instructions for Use for the 5cc and 10 cc bone void filler kits. No customer complaints or reports of injury have been received.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On June 7, 2019, customers were notified by letter sent by FedEx that Skeletal Kinetics LLC (Skeletal Kinetics) is voluntarily recalling certain OsteoVation, Inject Bone Void Filler kits distributed by OsteoMed LLC.The recall was initiated to correct a discrepancy between the Mixing Instruction Card and the Instructions for Use. The letter notes that Skeletal Kinetics has not received any customer complaints or reports of injury to the patient or end user. Customers are requested to: (1) review this letter in full, review current inventory, and complete and return the Acknowledgement and Receipt Form via email or the enclosed pre-paid envelope; and (2) return unused inventory via Federal Express with reference to Return Material Authorization FER 3003890476-6/6/19-001-R or by pickup by the sales representative indicated on the Acknowledgement and Receipt Form. Customers with questions are instructed to contact Mr. Randy Mosley, Director of QA, Skeletal Kinetics LLC by telephone at 972 677 4789 Monday-Friday, 8 AM 5 PM CDT.
Quantity in commerce
49 units
Distribution
U.S. Nationwide distribution including in the states of TX, NY, FL and AZ.
Product codes and lots
OsteoVation, Inject, 10cc (part number 390-2010, manufacturing lot 1130303, label no 12464). UDI (2): (01) 0 0813845 02006 1 (17) 210501 (10) 1130303

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0867-2021
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0866-2021Class IITerminatedVoluntary: Firm initiated
Z-0867-2021Class IITerminatedVoluntary: Firm initiated