Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 -
Beckman Coulter Inc. began a recall of Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used... on . FDA classified it as Class II and gives the reason as "Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1674-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Beckman Coulter MicroScan PROMPT Inoculation System-D, Catalog Number B1026-10D, UDI: 15099590659561 - Product Usage: The Prompt Inoculation System-D is used to standardize inocula for microdilution antimicrobial susceptibility test.
- Product code
- LIE Reagent/Device, Inoculum Calibration
- Reason for recall
- Due to customer reports it has been determined that the inoculation system contains the incorrect prompt wands. Use of the incorrect wands could result in false -susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 12/13/2019, the firm notified its customers via "Urgent Medical Device Recall" letter informing customers of the recall and its potential to cause false-susceptible results, false-resistant results, possible misidentifications, or possible out-of-range Quality Control results. The notification instructs customers to take the following actions: 1. Discontinue use and discard any remaining inventory of the incorrect wands with lot number 13.LAG.1 that may be found with the Prompt Inoculation System-D lot number 20201103T. 2. The notification letter also instructs customers to share the contents of the letter with lab/medical directors to determine the need to review previous patient and quality control test results. 3. In order to assure you have received the important communication, please respond within 10 days in one of the following ways: - Electronically, if you received this communication via email - Manually, complete and return the enclosed Response Form 4. If you have any questions regarding this notice, please contact the firm's Customer Support Center: - Via our website, http://www.beckmancoulter.com/customersupport/support - Via phone, call 1-800-677-7226 in the US 5. For customers in the US, if you need replacement products: - Complete the attached "Replacement Order Form" and email the form to [email protected] or fax to (866) 294-7850 or Call Client Services at (800) 526-3821
- Quantity in commerce
- 3,472 units
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MI, MS, NE, NY, NC, ND, OH, OK, PA, SD, TN, TX, UT, WV, WI, WY and countries of Canada, China, Japan, Paraguay.
- Product codes and lots
- Lot number 20201103T, Expiration Date: 11/3/2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1674-2020 | Class II | Terminated | Voluntary: Firm initiated |