Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate With Taper Adapter-indicated for primary...
Biomet, Inc. began a recall of Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate With Taper Adapter-indicated for primary, fracture, or revision total shoulder replacement for... on . FDA classified it as Class II and gives the reason as "Package is intended to contain one (1) baseplate and one (1) taper adapter; however, devices may be packaged with two (2) baseplates and no (0) adapters". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1615-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate With Taper Adapter-indicated for primary, fracture, or revision total shoulder replacement for the relief of pain and significant disability due to gross rotator cuff deficiency Item Number: 010000589
- Product code
- PHX Shoulder Prosthesis, Reverse Configuration
- Reason for recall
- Package is intended to contain one (1) baseplate and one (1) taper adapter; however, devices may be packaged with two (2) baseplates and no (0) adapters.
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Zimmer Biomet issued notification letter on 2/25/20 via email to the International consignees states reason for recall, health risk and action to take: Hospital Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certification of Acknowledgement Form and send to [email protected] form must be returned even if you do not have affected products available to return in your facility. 5. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please contact your Zimmer Biomet representative.
- Quantity in commerce
- 21 units
- Distribution
- International distribution - country of Netherland.s
- Product codes and lots
- Lot Number: 410420 UDI: (01) 0 0880304 53246 5 (17) 290825 (10) 410420
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1615-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.