Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1885-2020Class II

ACMI 9FR PROBE (E4-9F) Manufactured for Olympus Surgical Technologies, America (OSTA); NTI Catalog Number...

Northgate Technologies Inc. / initiated 2020-03-10 / Terminated / root cause: manufacturing process

Northgate Technologies, Inc. began a recall of ACMI 9FR PROBE (E4-9F) Manufactured for Olympus Surgical Technologies, America (OSTA); NTI Catalog Number 72-00198-0 - Product Usage: Probes are used for the... on . FDA classified it as Class II and gives the reason as "Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1885-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Northgate Technologies Inc.
Product
ACMI 9FR PROBE (E4-9F) Manufactured for Olympus Surgical Technologies, America (OSTA); NTI Catalog Number 72-00198-0 - Product Usage: Probes are used for the fragmentation of urinary/renal and biliary calculi.
Product code
FFK Lithotriptor, Electro-Hydraulic
Reason for recall
Investigation into a biocompatibility test discrepancy where a lot failed biocompatibility testing.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Northgate Technologies, Inc. notified customers on 03/11/2020 via "Urgent: medical Device Recall Notification" email. The recall notice identified the affected and requested the customers to cease distribution and destroy affected product. The customers were asked to complete the "Recall Acknowledgment Form".
Quantity in commerce
1238 boxes. 3 units per box.
Distribution
US Nationwide distribution. No governmental.
Product codes and lots
NTI Catalog Number 72-00198-0.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1885-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1883-2020Class IITerminatedVoluntary: Firm initiated
Z-1884-2020Class IITerminatedVoluntary: Firm initiated
Z-1885-2020Class IITerminatedVoluntary: Firm initiated