Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional...

Philips North America LLC / initiated 2020-03-05 / Terminated / root cause: design

Philips North America, LLC began a recall of Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the... on . FDA classified it as Class II and gives the reason as "No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1801-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm, Philips, sent an "URGENT-MEDICAL DEVICE CORRECTION" letter to the US and an "URGENT-Field Safety Notice" letter to the OUS (foreign countries) dated 2/26/20 on 3/5/2020 to all consignees by certified mail. These letters to all users explained that no torque was specified for the screws that connect the gearbox and the mounting flange within the C-arm roll motor assembly. A low torque could cause stress on the system and result in a loosening of the screws. If this occurs, the C-arm could rotate freely and uncontrolled movement could cause collision and/or entrapment. Philips asked that all users of the system be informed of the issues and to place a copy of the Field Safety Notice in the Instructions for Use for the system. Philips will exchange the roll motor in all affected systems free of charge, starting in March 2020. A local Philips representative will contact each consignee to schedule the roll motor replacement. If you need any further information or support concerning this issue, please contact your local Philips representative. Technical Support Line: 1-800-722-9377.
Quantity in commerce
162 units
Distribution
Worldwide distribution: US (nationwide) including the following states: Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin. And the following Countries of: Australia, Azerbaijan, Chile, Croatia, Czechoslovakia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Korea South, Lebanon, Mexico, Netherlands, Norway, Pakistan, Poland, Russian Federate, Senegal, Spain, South Africa, Sweden, Switzerland, Taiwan, and Uruguay.
Product codes and lots
Serial Numbers:  709, 710, 711, 713, 716, 715, 439, 526, 637, 653, 684, 621, 497, 792, 595, 674, 737, 628, 429, 605, 895, 540, 769, 668, 697, 758, 633, 520, 738, 691, 813, 428, 590, 706, 611, 655, 804, 663, 588, 877, 854, 654, 488, 836, 689, 717, 790, 629, 598, 736, 810, 841, 946, 795, 673, 776, 645, 646, 703, 608, 805, 746, 262, 730, 753, 878, 755, 837, 618, 594, 714, 543, 751, 263, 766, 817, 741, 534, 535, 536, 466  System Equipment Numbers: 74620447, 74620437, 74613625, 74613538, 74620341, 77839007, 73340583, 73326701, 77646555, 77559977, 77970948, 77427180, 72722066, 79132604, 72273457, 76100048, 77970800, 74620302, 72805859, 77228623, 80018072, 74611961, 77647014, 75486269, 78248928, 75486357, 76888124, 75581571, 78243622, 75423992, 77971015, 74817549, 76890243, 74352012, 77260210, 77213972, 78993356, 77347585, 75532686, 79152378, 79109118, 75533239, 74536213, 79152254, 77838972, 78150020, 78696158, 77559984, 76481840, 75848181, 79112403, 77905831, 80171865, 79042710, 77697096, 77213970, 74778339, 74778897, 73051857, 73717291, 78132484, 75305793, 72312081, 76890941, 76888973, 78365267, 76888978, 79464469, 72323525, 72323935, 74169336, 74620425, 76891009, 72417922, 79035318, 79035262, 78367631, 74584486, 74584491, 75615010, 75470091

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1801-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1801-2020Class IITerminatedVoluntary: Firm initiated