Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1763-2020Class II

Olympus Pleuravideoscope, Model LTF-240

OLYMPUS Corporation of the Americas / initiated 2020-03-23 / Terminated

Olympus Corporation of the Americas began a recall of Olympus Pleuravideoscope, Model LTF-240 on . FDA classified it as Class II and gives the reason as "There was no data to validate that the identified endo-therapy devices were compatible with certain pleuravideoscope models. Updated Operations Manuals are being provided". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1763-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus Pleuravideoscope, Model LTF-240
Product code
GCJ Laparoscope, General & Plastic Surgery
Reason for recall
There was no data to validate that the identified endo-therapy devices were compatible with certain pleuravideoscope models. Updated Operations Manuals are being provided.
FDA root cause
Under Investigation by firm
Action
An Urgent Medical Device Correction notification letter dated 3/23/20 was sent to customers. Actions to be taken by the user: Our records indicate that your facility has purchased one or more of the above-referenced endoscopes. OLYMPUS requires that you take the following actions: a) Inspect your inventory of operation manuals for the referenced Olympus models and discard the existing ones. b) Exchange them with the updated, enclosed Operation Manuals having an updated system chart indicating compatible endoscope accessories. c) Notify related personnel at your institution that they should stop the combination use of the LTF-160 and LTF-240 and the subject Endo-Therapy accessories. d) Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0066 and provide your contact information as indicated in the portal. e) If you have further distributed this product, identify your customers, forward them this notification including the attachments, and appropriately document your notification process. Olympus regrets any inconvenience caused and fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly at (484)896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at [email protected].
Quantity in commerce
265 total
Distribution
US Nationwide
Product codes and lots
All Serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1763-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1762-2020Class IITerminatedVoluntary: Firm initiated
Z-1763-2020Class IITerminatedVoluntary: Firm initiated