510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1692-2020Class II

djosurgical Linear Neck Trials with Spring Sides, Standard, nonsterile, REF S-200635; and djosurgical Linear...

Encore Medical, LP / initiated 2018-07-10 / Open, Classified / root cause: manufacturing process

Encore Medical, LP began a recall of djosurgical Linear Neck Trials with Spring Sides, Standard, nonsterile, REF S-200635; and djosurgical Linear Neck Trials with Spring Sides, Lateral... on . FDA classified it as Class II and gives the reason as "The retaining ring on the neck trials has a potential to fail interoperatively". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1692-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
djosurgical Linear Neck Trials with Spring Sides, Standard, nonsterile, REF S-200635; and djosurgical Linear Neck Trials with Spring Sides, Lateral, nonsterile, REF S-200636. The firm name on the label is djosurgical, Austin, TX.
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
The retaining ring on the neck trials has a potential to fail interoperatively.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The recalling firm issued a letter dated 7/10/2018 to their field representatives (agencies) via email explaining the reason for recall.
Quantity in commerce
28 (14 REF S200635 and 14 S200636)
Distribution
Distribution was made to CA, CO, MD, MN, OH, TX, and WA. There was no foreign/military/government distribution.
Product codes and lots
All Revision C trials. REF S-2000635: 313R1000, 313R1900, 313R1901; REF S-200636: 314R1000, 314R1900, 314R1901.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1692-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1692-2020Class IIOngoingVoluntary: Firm initiated