The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical...
Biomerieux Inc began a recall of The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in-vitro test to determine... on . FDA classified it as Class II and gives the reason as "There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card (ref. 421857, lot 8021232403)". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1742-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- bioMerieux, Inc.
- Product
- The VITEK 2 Gram-positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in-vitro test to determine susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed.
- Product code
- N/A
- Reason for recall
- There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card (ref. 421857, lot 8021232403). .
- FDA root cause
- Under Investigation by firm
- Action
- On March, 11, 2020 a notification letter was issued to the impacted subsidiaries which included a Customer Letter to be translated into local language. The customer letter instructs the customer to stop the use of the impacted AST-P652 lot, to complete the attached Acknowledgement Form, and return any affected unused product to their local bioMrieux representative as soon as possible.
- Quantity in commerce
- 1698 cartons (15,940 cards)
- Distribution
- No distribution within the United States. Foreign distribution locations: Netherlands, Luxembourg, Belgium
- Product codes and lots
- VITEK 2AST-P652 Test Kit with Catalog #421857 and Lot# 8021232403 Manufactured 11OCT2019, Expires 11APR2021
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1742-2020 | Class II | Ongoing | Voluntary: Firm initiated |