NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI...
Nobel Biocare Usa Llc began a recall of NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are... on . FDA classified it as Class II and gives the reason as "The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1851-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nobel Biocare USA LLC
- Product
- NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI (01)07332747091441(10)12129104(17)240203 - Product Usage: are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- The top label of article no. 37965, lot 12129104 (immediate packaging of endosseous dental implants ) may incorrectly indicate an implant length of 7mm, which is shorter than the actual length of the implant (10mm).
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- The firm's "Urgent Field Safety Notice", was issued on 08/28/2019 and the notices were sent to customers via email . The following actions are to be taken: 1) Identify affected devices in stock by confirming the information on the side label: 3.75x10mm, article no. 37965, lot 12129104. 2) Confirm that both the side and top label indicate 3.75x10mm. 3) Complete a Customer Acknowledgment Form and return it via email to: [email protected] within 5 days of receipt of the notice. 4) Return incorrectly labeled products (top label 3.75x7mm) to: Attn: Returns PFA1925 109004 Moscow Stanislavskogo str., h.21, b.3 (Correctly labeled products (top label 3.75x 10mm) do not need to be returned). 5) Returned devices will be replaced free of charge.
- Quantity in commerce
- 228 pieces
- Distribution
- International distribution - the country Russia only.
- Product codes and lots
- Lot:12129104
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1851-2020 | Class II | Terminated | Voluntary: Firm initiated |