Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human...
CSL Behring GmbH began a recall of Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997 on . FDA classified it as Class II and gives the reason as "Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1790-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CSL Behring LLC
- Product
- Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997
- Product code
- NAO Test, Carbohydrate Deficient Transferrin
- Reason for recall
- Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring
- FDA root cause
- Under Investigation by firm
- Action
- Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-20-002.A.US to the 5 affected customers in the U.S. on March 5, 2020 via FedEx overnight delivery. ( Urgent Field Corrective Action (UFCA) PP20-002.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities). Letter states reason for recall, health risk, and action to take by the Customer Please review this letter with your Medical Director. Discontinue use of and discard the kit lots listed in Table 1. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
- Quantity in commerce
- 297 PCK US; 1247 PCK OUS
- Distribution
- IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan
- Product codes and lots
- US Lot number: 49822 UDI: 842768018534 OUS: Lot numbers: 49729, 49849, 49985, 50084 Exp. Date: 2020-04-11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1790-2020 | Class II | Terminated | Voluntary: Firm initiated |