PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
Merit Medical Systems, Inc. began a recall of PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT on . FDA classified it as Class II and gives the reason as "During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1796-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- PreludeSYNC DISTAL Radial Compression Devices, SDRB-REG-LT, SDRB-REG-RT
- Product code
- DXC Clamp, Vascular
- Reason for recall
- During the sterilization of radial compression devices, some of the inflatable balloons may break or burst, which may result in air leakage and the inability to maintain pressure/compression. This may result in a lack of hemostasis resulting in bleeding or a delay in treatment; the defect is likely to be visually identified by the clinician before use or identified during inflation of the balloon.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- On 03/11/2020, Urgent Product Recall Notices were mailed via second day service. The recalling firm requests that customers immediately stop using, quarantine, and return the affected lots. Further, they requested that customers make personnel within their organization aware of this field action. If the product was further distributed to other facilities, institutions, or manufacturers, customers were asked to ensure this notice was immediately shared with them and note the quantity distributed on the customer response form. Customers with additional questions were encouraged to call 801-208-4381.
- Quantity in commerce
- 1932
- Distribution
- U.S.: NV, IL, VA, PA, CA, AR, MA, GA, OH, NJ, TX, FL, NC, SC, MI, NH, CO, AL, MN, KS, NM, AZ, ID, MD, NY, OR, DC, DE, MO, IN, WI, IA, TN. O.U.S.: Switzerland, Hong Kong, Spain, Portugal, Canada, Singapore, United Arab Emirates, Mexico, United Kingdom, France, Japan, Germany, Italy
- Product codes and lots
- SDRB-REG-LT: Lot(UDI): H1559517(10884450363673) and H1727668(00884450363676). SDRB-REG-RT: Lot(UDI): H1666548(10884450363659) and H1723973(10884450363659)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1796-2020 | Class II | Terminated | Voluntary: Firm initiated |