Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1835-2020Class II

ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499

Abbott Medical / initiated 2019-07-12 / Terminated / root cause: software

ICU Medical, Inc. began a recall of ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499 on . FDA classified it as Class II and gives the reason as "Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1835-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499
Product code
DXQ Blood Pressure Cuff
Reason for recall
Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary parameter.
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
Customers were notified of the recall via UPS letter on 7/12/2019. The firm is voluntarily notifying customers and will contact customers to schedule a software update once the updated software is available. The firm has instructed customers to use the mitigation instructions detailed in the customer letter to reduce the likelihood of the software conditions occurring. The firm also requested customers complete a response form and return it to the firm. For additional information or technical assistance contact ICU Medical at 1-866-829-9025, option 4.
Quantity in commerce
83 units
Distribution
US Nationwide Distribution in the states of FL, WI, NY, NC, VT, IL, MD, CA, NE, WA OUS: None
Product codes and lots
All serial numbers of Cogent HMS, PN 58400-000, PN 58400-000R (User Interface Module (UIM), PN 58403-000 and Patient Interface Module (PIM), PN 58401-000) with Software Versions 1.3.2.460 and prior are impacted.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1835-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1835-2020Class IITerminatedVoluntary: Firm initiated