ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499
ICU Medical, Inc. began a recall of ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499 on . FDA classified it as Class II and gives the reason as "Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1835-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- ICU Medical Cogent Hemodynamic Monitoring System, Cogent Item Number 58400-000, UDI: (01)00840619079499
- Product code
- DXQ Blood Pressure Cuff
- Reason for recall
- Firm identified several software issues through internal testing of its Cogent Hemodynamic Monitoring System. These issues may result in inaccurate monitoring data, unit shutdown, and/or loss of display of a secondary parameter.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- Customers were notified of the recall via UPS letter on 7/12/2019. The firm is voluntarily notifying customers and will contact customers to schedule a software update once the updated software is available. The firm has instructed customers to use the mitigation instructions detailed in the customer letter to reduce the likelihood of the software conditions occurring. The firm also requested customers complete a response form and return it to the firm. For additional information or technical assistance contact ICU Medical at 1-866-829-9025, option 4.
- Quantity in commerce
- 83 units
- Distribution
- US Nationwide Distribution in the states of FL, WI, NY, NC, VT, IL, MD, CA, NE, WA OUS: None
- Product codes and lots
- All serial numbers of Cogent HMS, PN 58400-000, PN 58400-000R (User Interface Module (UIM), PN 58403-000 and Patient Interface Module (PIM), PN 58401-000) with Software Versions 1.3.2.460 and prior are impacted.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1835-2020 | Class II | Terminated | Voluntary: Firm initiated |