Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1748-2020Class II

Fresenius 2008T Hemodialysis Machine: Part Number Description 190573 2008T HEMODIAL YSIS SYS. OLC/DIASAFE...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2020-03-12 / Open, Classified / root cause: software

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius 2008T Hemodialysis Machine: Part Number Description 190573 2008T HEMODIAL YSIS SYS. OLC/DIASAFE PLUS; 190713 2008T HEMODIAL YSIS SYS., with CDX... on . FDA classified it as Class II and gives the reason as "A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machine". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1748-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius 2008T Hemodialysis Machine: Part Number Description 190573 2008T HEMODIAL YSIS SYS. OLC/DIASAFE PLUS; 190713 2008T HEMODIAL YSIS SYS., with CDX; 190766 2008T HEMODIAL YSIS SYSTEM W/BIBAG; 190858 2008T HEMODIAL YSIS SYSTEM WITHOUT CDX; 190895 2008T GEN 2 BIBAG WITHOUT CDX; 190908 2008T HD System With CDX, CAN
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machine
Root cause tag
software
FDA root cause
Software design
Action
Fresenius Medical Care issued via First Class Mail on 03-12-2020 Important Medical Device Correction Customer Notification explaining the issue, health risk and action to take: To clear the alarm, tum the machine power off for 30 to 45 seconds and turn back on. If the alarm has not been cleared, repeat the power off cycle again. If still not cleared, perform a manual blood return and remove the patient from the machine. Remove the machine from service and contact a qualified service technician. This alarm may reoccur even after a power cycle has been performed. true alarm will only occur when an externally powered USB device is connected. Refer to the Operators Manual Troubleshooting section for instructions. A Reply Form included for confirmation the consignee read and understands the notice sent to them and ask for number of machines corrected. A software solution is being created that addresses this situation. Once the software upgrade is available, an additional notification will be sent with instructions on how to order the upgrade at no charge. If you experience this issue or have additional questions, please contact Fresenius Medical Care Technical Services at 800-227-2572
Quantity in commerce
101,463
Distribution
Nationwide Foreign: Canada, Curaca, Guam
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1748-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1748-2020Class IIOngoingVoluntary: Firm initiated
Z-1749-2020Class IIOngoingVoluntary: Firm initiated