Z-2131-2020Class II

da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery

Intuitive Surgical, Inc. / initiated 2019-08-19 / Open, Classified / root cause: manufacturing process

Intuitive Surgical, Inc. began a recall of da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery on . FDA classified it as Class II and gives the reason as "Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2131-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Urgent Medical Device Recall Letters (Field Safety Notices) were hand-delivered to U.S. customers on August 19, 2019 and OUS customers on August 20, 2019. Customers were asked to locate affected camera(s) in their inventory and inspect per the instructions attached to the Field Safety Notice throughout the life of the camera(s). If residual soil is present, customers were to return the affected camera(s) to the manufacturer using the standard Return Materials Authorization (RMA) process. Additionally, customers were requested to: iInform necessary hospital personnel about this issue; complete the attached Acknowledgement Form and return it via fax to Intuitive as instructed on the form; and to retain a copy of this letter and the Acknowledgement Form for their files. Customers were informed that an Intuitive Representative would schedule a site visit to review the attached inspection instructions, and that the firm will provide a replacement camera, at no charge, for returned affected product. Customers requiring further information or support may contact their Clinical Sales Representative or contact Intuitive Customer Service at the following numbers: US Customer Service (800) 876-1310, Option 3 (7 AM to 5 PM EST) or [email protected], or South Korea Customer Service: + 82-02-3271-3200, Option 3 ( 9AM to 6 PM KSTJ)
Quantity in commerce
305 units
Distribution
Distribution US nationwide and South Korea.
Product codes and lots
Model Number: 430060-33; UDI Number: 00886874114384; All serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2131-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2131-2020Class IIOngoingVoluntary: Firm initiated