Z-1789-2020Class II

Temperature Sensing 100% Silicone Foley Catheter, Part Number 102203101863DO -

Degania Medical Devices Pvt. Ltd. / initiated 2020-03-17 / Terminated / root cause: manufacturing process

Degania Medical Devices Pvt. Ltd. began a recall of Temperature Sensing 100% Silicone Foley Catheter, Part Number 102203101863DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure... on . FDA classified it as Class II and gives the reason as "Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1789-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Degania Medical Devices Pvt. Ltd.
Product
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102203101863DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Product code
EZL Catheter, Retention Type, Balloon
Reason for recall
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An Urgent Field Safety Notice dated 3/17/20 was sent to customers. Please immediately follow the Required Customer Actions set fo rth below. 1. Please forward this information to the immediate attention of the persons responsible for usage of the affected parts. Sub-recall must be initiated by your company to collect any distributed product from the suspected list. 2. Please fill in the attached Recall Response form and indicate the quantity of the affected product in your possession. 3. Please promptly return the completed Recall Response fo rm to Degania Silicone by e-mail: [email protected]. 4. The disposition of the recalled goods should be scrap. We would ask you to destruct the goods at your facil ity and send us the Scrap confirmation letter. Should you have any questions about this subject, please contact Degania Silicone: Degonia Silicone Ltd. Degania Bet, 1513000, ISRAEL Katia Kayam, Cust omer compla ints Officer Tel: +972-529532059 E-mail: [email protected]
Quantity in commerce
1578
Distribution
Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.
Product codes and lots
Lot Numbers: V19008709 V19016709

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1789-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1779-2020Class IITerminatedVoluntary: Firm initiated
Z-1780-2020Class IITerminatedVoluntary: Firm initiated
Z-1781-2020Class IITerminatedVoluntary: Firm initiated
Z-1782-2020Class IITerminatedVoluntary: Firm initiated
Z-1783-2020Class IITerminatedVoluntary: Firm initiated
Z-1784-2020Class IITerminatedVoluntary: Firm initiated
Z-1785-2020Class IITerminatedVoluntary: Firm initiated
Z-1786-2020Class IITerminatedVoluntary: Firm initiated
Z-1787-2020Class IITerminatedVoluntary: Firm initiated
Z-1788-2020Class IITerminatedVoluntary: Firm initiated
Z-1789-2020Class IITerminatedVoluntary: Firm initiated