Z-1689-2020Class II
Jamshidi Bone Marrow Biopsy/Aspiration Needle, with Luer Lock Adapter 11GX4 ASP, Sterile -
Bard Peripheral Vascular Inc / initiated 2020-03-17 / Terminated / root cause: packaging
Bard Peripheral Vascular Inc began a recall of Jamshidi Bone Marrow Biopsy/Aspiration Needle, with Luer Lock Adapter 11GX4 ASP, Sterile - Product Code: DJ4011X - Product Usage: The device is comprised of an... on . FDA classified it as Class II and gives the reason as "Sterility compromised due to packaging pouches that may not be sealed properly". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1689-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- Jamshidi Bone Marrow Biopsy/Aspiration Needle, with Luer Lock Adapter 11GX4 ASP, Sterile - Product Code: DJ4011X - Product Usage: The device is comprised of an outer cannula with a handle and an inner stylet. Its intended use is for the posterior iliac crest biopsy technique.
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- Sterility compromised due to packaging pouches that may not be sealed properly
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Becton, Dickinson and Company (BD) issued notification on 3/17/20 stating reason for recall, health risk, and action to take: check all inventory locations within your institution for affected product code / lot number combinations listed in the recall notice. If you have further distributed any of the product code / lot numbers, immediately contact that location, advise them of the recall, forward these instructions and have them complete the instructions within this recall notice. 3. Please remove any identified product from your shelves. 4. Destroy all identified product from the affected product code / lot number combinations following your institutions process for destruction. 5. If you have used the affected product, complete and return the attached Recall and Effectiveness Check Form indicating that you do not have affected product in inventory. Once the product affected by this recall has been removed from your inventory and destroyed: 6. Fill out the Recall and Effectiveness Check Form. Be sure to state the quantities and lot numbers of each recalled product that you destroyed. It is extremely important that we receive this information. 7. Email the completed Recall and Effectiveness Check Form to BD Regulatory Compliance at [email protected] or fax it to 312-949-0229. 8. Once the Recall and Effectiveness Check Form has been completed and emailed to BD Regulatory Compliance at [email protected] or faxed to 312-949-0229 and all information has been verified, BD will issue you a product replacement for any destroyed inventory. Please call our BD Customer/Technical Support at 800-638-8663 with any questions.
- Quantity in commerce
- 8,170 units (US: 420 and OUS: 7,750)
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of FL, IL, NJ, NY, OH, TN, VA and countries of AUSTRALIA, INDIA, JAPAN, MALAYSIA, NEW ZEALAND, PHILIPPINES and SINGAPORE.
- Product codes and lots
- Lot Numbers: 0001303256 and 0001303257
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1689-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY